Description
Philips Veradius Unity Flat Detector C-Arm
FEATURES:
- Wireless foot switch.
- Accurate repositioning with Position Memory.
- Dose Aware real-time dose feedback.
- Less than 70 seconds system startup.
- ClearGuide.
- Incredible image quality.
- Dual 19” color LCD monitors.
- Flat Detector
Fast Positioning
Philip’s ClearGuide and color coding on the Veradius, speed up positioning. ClearGuide provides a set of reference number for the operator to use during positioning.
Position Memory
The Philips C-Arm can store up to four positions, and can recall and display the current position and and a saved position on the c-arm stand monitor.
Flat Detector
The Veradius has a third generation flat detector. It provides high-quality fluoroscopy, exposure, subtraction runs, and roadmap guidance to support orthopedic, cardiac, and vascular surgeons on performing the most challenging procedures. In the idle state, the flat detector calibrates itself every 5 minutes. It has a field of view of 10.3″ and a grid ratio of 13:1. It can be easily removed by manually unscrewing four bolts. This allows for great image quality with X-ray dose efficiency for imaging small anatomies. The high dynamic range of the flat detector (96 dB) provides higher contrast images for demanding procedures. With the ability to differentiate a larger range of grey values than image intensifiers.
C-Arm Movements:
Angulation: 140° rotation (+90/-50°) offers a maximum of projection flexibility
Motorized hight movement: 19.3” (49 cm)
Longitudinal Movement: 7.0” (20 cm)
Panning Movement (Swivel): ±10°
Rotation: ±200°
Lowest Lateral Position: 40.4” (201.7 cm)
Source to Image Distance (SID) 39.1” (99.3 cm)
Free Space in C-Arm: 30.4” (77.2 cm)
C-Arm Depth: 29” (73 cm)
Parallel Movement: Dedicated parallel movement with rear-wheel steering, for easy positioning along operating table.
C-arm stand length: 77.2” (196 cm)
Mobile View Station Monitors:
Resolution: 1280 x 1024 pixels
Maximum light output: 650 cd/m2
Minimal Contrast Ratio: 1/400
Contrast Ratio: >700:1
Viewing Angle: 170° in horizontal and vertical direction.
Two Touchscreen 19” high-brightness color LCD monitors
X-Ray Generator:
Monoblock
High tension transformer
Frequency: 80 kHz
X-Ray Tube:
Rotating Anode
High-Power geneartor
X-Ray:
Tube Type: Rotating Anode
Nominal X-ray tube voltage: 120 kV
Nominal focal spot values: 0.3 and 0.6 mm
Max. Anode Heat Content: 315 kHU
Max. Heat Dissipation: 73 kHU/min
Cooling Method: Active oil-circulation cooling
Inherent filter: 1.0 mm AI. equivalent
Integrated Beam Filter: 3.0 mm Ai +0.1mm Cu
Total Beam Filtration: 6.75 mm AI. eq
Operating Values With Pulse Fluoroscopy:
kV range: 40-120 kV
mA Peak Range: 1-60 mA
Pulse Width: 5.6-44 ms
Pulse Rate: 4-8-15-optional 30
Operating Values With Exposure Runs:
kV Range: 40-120 kV
mA Peak Range: 1-60 mA
Pulse Width: 9.3 178 ms
Pulse Rate: 4-8-15-optional 1-2-30
Opearting Values With Single Shot Exposure
kV Range: 40-120 kV:
mA Peak Range: 2-125 mA
Time Range: 100 and 165 ms
X-Ray Collimation:
Full lead shutters can be rotated and moved independently.
Automatic Shutter Positioning (ASP) automatically positions shutters for optimal image quality.
Flat Detector:
Flat Detector Frees up valuable workspace around the patient during challenging procedures.
Integrated Laser in the flat detector:
635 nm
Max. Output: <5 mW
Communications:
IP Addressing: Static IP, DHCP
Wireless Standards Supported: IEEE802.11a/b/g
User Configurable SSID Support:Up to 16 SSID, each with a uniqueMAC address and configurable SSIDBroadcast.
Number of Roaming Points Supported:16
Authentication Protocols: IEEE802.1xEAP-SIM, EAP-TLS, EAP-TTLS, EAP-FAST and PEAP (With choice of AES,TKIP and WEP encryption)
Security: IEEE802.11i WPA2, WPA withAES or TKIP encryption.
Flat Detector not included in warranty
Terms: 50% Due at time of Order, 40% Prior to ship, 10% NET 15