Steris AMSCO Century Small Sterilizer – Steris SV-120 Steam Sterilizer

Request Quote

The Steris AMSCO SV-120 Small Steam Sterilizer is a high-performance, pre-vacuum sterilizer designed for reliable sterilization of nonporous, heat- and moisture-stable materials. Featuring a spacious 20″ x 20″ x 38″ (250 liters / 66 gallons) chamber made from durable 316L stainless steel, it is ideal for sterilizing liquids, media, glassware, instruments, clothing, and laboratory waste. Equipped with vertical doors with glass panels, an ALUS chamber rail for racks, and a 6.5-inch high-resolution color touch screen, the sterilizer offers intuitive operation and comprehensive cycle tracking via external USB, PDF reports, and a built-in matrix printer. Additional features include independent temperature probes, pneumatic valves, safety controls, and a water-saving system for the vacuum pump. Certified for safety and performance, the SV-120 ensures consistent, repeatable sterilization for demanding laboratory and medical applications.

To submit a quote request for this product please fill out the form below.
Product Quote Request
Name
Name
Quantity *

Description

Steris AMSCO Century Small Sterilizer – Steris SV-120 Steam Sterilizer

Steris AMSCO SV-120 Prevacuum Steam Sterilizer (20″ x 20″ x 38″ Chamber, 250 Liters / 66 gallons) is designed for sterilization of nonporous heat- and moisture-stable goods, sterilization of liquids and media applications. This laboratory steam sterilizer provides high quality repeatable performance and accountability for a wide range of lab applications, including: liquids, media, glassware, instruments, clothing, waste, media clothing and laboratory items. 

Features:

  • Chamber material and pipe work 316L stainless steel
  • Fully jacketed chamber
  • Vertical opening doors with glass panel
  • ALUS chamber rail for racks
  • 6.5” (165 mm) high resolution color touch screen
  • Interface to tracking system XLM and PDF format
  • External USB connection for data download (on loading side)
  • Matrix printer on loading side with A4 print out capability
  • Additional touch screen provided on the unloading side
  • Independent temperature probe (chamber)
  • Pneumatic valves
  • Air filter
  • Pressure gauges for jacket, chamber, steam generator (loading side), chamber (unloading side)
  • Water level gauge for steam generator
  • Safety temperature switch for door opening
  • Acoustic signal for end cycle and alarms
  • Programmable timer start for test cycles
  • Potential free outputs for critical alarms
  • Sanitary piping with Tri-clamp
  • Water Saving System For Vacuum Pump
  • Separate water feed line for steam generator and vacuum pump
  • Drain discharge cool down

Standards:

  • Each sterilizer meets applicable requirements of the following listings and standards, and carries the appropriate symbols:
    • Underwriters Laboratories (UL) Standard 544 as certified by ETL Testing Laboratories, Inc.
    • Canadian Standards Association (CSA) Standard C22.2 No. 125.
    • ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form R-1 is furnished. Shell and door are designed for maximum allowable working pressure of 42 psig (289.6 kPa) for 16 x 16 x 26″(406 x 406 x 660 mm) sterilizers and 40 psig (275.8 kPa) for 20 x 20 x 38″ (508 x 508 x 965 mm) sterilizers.
    • ASME Code, Section I, Part PMB, for power boilers, if optional steam generator is supplied.
    • Seismic Pre-approval R-0282.

Breakdown of Prevacuum vs. Gravity 

  • Gravity - designed for sterilization of liquids in flasks with vented closures, at 250°F (121°C), and also heat- and moisture-stabile goods at 270°F (132°C).
  • Prevacuum – designed for efficient, high-volume processing of heat- and moisture-stabile materials, such as fabrics, wrapped hard goods, glassware or animal bedding, at 270°F (132°C). Prevacuum units will also run gravity-type cycles.

Gravity units operate by the gravity downward displacement air-removal principle. Prevacuum sterilizers have a chamber evacuation system for high-speed, high-volume sterilization. Both types use steam-under-pressure as the sterilizing agent.