Zimmer ATS 750

Request Quote

The Zimmer A.T.S. 750 Automatic Tourniquet System is a compact, single-port pneumatic surgical tourniquet designed to provide precise, microprocessor-controlled cuff pressure for extremity procedures. Operating on ambient air with no nitrogen tank required, it offers an adjustable pressure range of 50–475 mmHg in 5 mmHg increments, automatic pressure regulation, large red LED displays for pressure and time, and audible/visual alarms for key safety conditions such as high or low cuff pressure, leaks, time exceeded, and line occlusion. The unit includes a rechargeable battery backup providing at least 30 minutes of operation, supports tabletop or IV-pole mounting, and can be used with an optional dual control valve accessory for dual-cuff applications. Compact and portable, the A.T.S. 750 is well suited for operating rooms and surgical environments requiring reliable, easy-to-use tourniquet control.

To submit a quote request for this product please fill out the form below.
Product Quote Request
Name
Name
Quantity *

SKU: AM-ATS750 Categories: , Tags: , Brand:

Description

Zimmer ATS 750

The Zimmer A.T.S. 750 is a single-port automatic pneumatic surgical tourniquet system intended for qualified medical professionals to temporarily occlude blood flow in a patient’s extremities during extremity surgery.

Key features

  • Uses ambient air — no nitrogen tank required.
  • Electronic / microprocessor-regulated pressure for automatic pressure control.
  • Single-port system designed for single-port, single-bladder cuffs.
  • 30-minute battery backup for short-term mobility/backup operation.
  • Audible and visual alarms for unsafe or fault conditions.
  • Self-check calibration at power-up.
  • Large red LED displays for pressure and time, plus a green AC power indicator.
  • Tabletop or IV-pole mounting capability.
  • Dual-cuff capability only with an optional Dual Control Valve accessory for bilateral/Bier Block/IVRA use.

Main specifications

Category Spec
Pressure range 50–475 mmHg, adjustable in 5 mmHg increments
Pressure accuracy ±5 mmHg over 50–475 mmHg
Pressure regulation ±6 mmHg of set point, 10-second average, under non-transient conditions without external leaks
Maximum cuff pressure 475 mmHg
Time alarm range 5–240 minutes, 5-minute increments per service manual
Timer accuracy 0.25% of elapsed time
Power input 100–240 VAC, 50/60 Hz, auto-switching
Line current 670 mA RMS @ 120 V
Input power 53 W typical
Battery Rechargeable 12 V sealed lead-acid, 1.2 Ah
Battery run time Minimum 30 minutes with fully charged batteries
Battery recharge time 24 hours
Power cord Type SJT, AWG 16, 14 ft / 4.27 m, hospital-grade 3-prong plug
Fuses Two time-delay 1.0 A, 250 V fuses
Dimensions 8 in H × 7 in W × 7.9 in D / 20.3 × 17.8 × 20.1 cm
Weight 7.7 lb / 3.5 kg;
Displays Pressure: red 14-segment LED; Time: red 14-segment LED; AC indicator: green LED
Controls On/Standby, rotary knob, Pressure, Time, Inflate, Deflate, Alarm Silence
Safety classification UL 60601-1 Class I and internally powered, Type B, continuous operation; not for use with flammable anesthetic or gases
Internal diagnostics Program, memory, watchdog timer, transducer calibration, improper valve actuation

 

Alarm / safety functions

The system provides alarms for low cuff pressure, high cuff pressure, cuff-side leak, time exceeded, cuff inflated on power-up, cuff not deflated, and line occlusion. Low/high pressure alarms trigger at about 15 mmHg below/above set point, and leak alarm indicates a substantial leak lasting more than 7 seconds. Most audible alarms can be silenced for 30 seconds using the Alarm Silence switch.