Description
The LSP AutoVent series integrates field experience and engineering excellence, producing a lightweight, transportable automatic ventilator.
second-version 2000
Control Modules:
Weight: 24 oz. (680 grams).
Dimensions: 6″ (W) x 1.75″ (D) x 3.5″ (H) (15.24 cm x 4.4 cm x 8.9 cm).
Supply Pressure Range: 40 – 60 psig.
Operating Temperature: 0° F to 125° F.
Storage Temperature: -40° F to 160° F.
Frequency – AV2000: 8 to 20 BPM.
Tidal Volume – AV2000: 400 to 1200 ml.
Flow Rate: 16 to 48 lpm.
Inspiratory Time – AV2000: approx. 1.5 or 2.0 seconds.
Expiratory Time: 1.5 to 6.0 seconds I-E Ratio: 1:1 to 1:4.
Case Material: Polyester.
Input connection: Plated brass.
Output Connection: Plated brass minute volume: 0 to 24 lpm.
Gas Consumption Driving Gas: 0.5 lpm maximum.
Driving Gas: 0.5 lpm maximum.
Patient Valve Assembly:
Weight: 16 oz. (450 g).
Flow: 0 – 48 lpm at 50 psig.
Peak Inspiratory Flow: 48 lpm at an airway pressure drop of less than 2.5 H2O.
Delivery Pressure: 60 ± 5 cm H2O.
(Insp.) Crack Pressure: 0 to -2 cm (-0.8”) H2O.
Exhalation Resistance:
0 – 10 lpm: 1.5 cm H2O (max).
11 – 70 lpm: 3.8 cm H2O (ma