Description
Olympus CLV-190 Evis Exera II Light Source Features
Main Features:
NBI (Narrow Band Imaging) in EVIS EXERA III 190 Series scopes provides twice the viewable distance of EVIS EXERA II 180 Series scopes and offers significantly brighter images.
CV-190 features the ability to switch the point of focus between ‘near’ and ‘normal’ with the push of a button.
The newly designed, waterproof one-touch connector enables a one-step connection to the light source and does not require a separate scope cable for the video processor.
New and improved image processing delivers sophisticated image quality via enhanced color reproduction, minimized image noise, and reduced halation.
The pre-freeze function selects the clearest still image automatically, saving time.
The CV-190 is compatible with EVIS 100/130/140/150 Series, EVIS EXERA 160 Series, EVIS EVERA II 180 Series, GI/BF/ VISERA Series scopes, and EVIS EXERA III 190 Series.
16:9 and 16:10 outputs for a HDTV monitor are available. The system is compatible analog, HD-SDI, and DVI output.
A link connection to peripheral devices avoids complicated cable connections and accelerates transmission speed.
The Olympus documentation system enhances networking expandability.
Picture-in-picture and index functions effectively enhance your observation.
The system is compatible with portable memory, which is standard for data management; simply connect and upload.
The CV-190 supports DV output to compatible documentation devices
CLV-190 Features:
Examination Lamp: Xenon short-arc lamp (ozone-free) 300 W
Average Lamp Life: Approximately 500 hours of continuous use (With intermittent use, the lamp life may vary slightly.)
Ignition Method: Switching regulator
Brightness Adjustment: Light-path diaphragm control
Cooling: Forced-air cooling
Intensity Mode: Normal or high intensity
NBI Observation: Available
Color Conversion: Possible using special-purpose filter
Emergency Lamp: Halogen lamp (within mirror) 12 V 35 W
Average Emergency Lamp Life: Approximately 500 hours
Automatic Brightness Adjustment
Automatic Brightness Adjustment Method: Servo-diaphragm method
Automatic Exposure: 17 steps
Air Feeding
Pump: Diaphragm-type pump
Pressure Switching: 4 levels available (off, low, medium, high)
Water Feeding
Method: Feeds water by pressurizing the detachable water container with air
Indicators on Front Panel
Emergency Lamp: Indicates absence of emergency lamp, disconnection, and use of the emergency lamp
NBI: When NBI observation is enabled, the NBI indicator lights up
Setting Memory
Settings (except filter setting) are stored even when the light source is off.
Includes keyboard pigtail, water bottle