Description
Steris V148 Steam Sterilizers are designed for fast, efficient sterilization of heat and water-resistant materials with the same capabilities as a gravity sterilizer. The V148 pre-vacuum sterilizer is equipped with pre-vac, gravity, flash, express, leak test, and daily air removal test cycles. With a 26″ x 37.5″ x 48″ chamber, the V148 sterilizer provides enough space fit all of your equipment and sterilize with speed and reputable quality in mind.
Features:
Florescent display for easy reading
Ink on paper impact printer
26″ x 37.5″ x 48″ chamber
Pre-vacuum Steam Sterilizer
Fast & efficient
Emergency manual exhaust valve
Non-lubricated, steam-activated door seal
Help screens for programming/troubleshooting alarm conditions
Communication interface compatibility w/ most PC devices
Electronic water conservation valve
Optional Stainless steel side panels available.
Additional Options: Cabinet Package, Loading Cart, Integral Steam Generator, Water Filtration Options
Prevacuum configuration sterilizers are factory programmed with the following cycles:
270°F (132°C) Prevacuum (with either 5-minute or 20-minute dry phases) – For efficient, high volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 5 minutes or 20 minutes
Gravity, Liquid, 250°F (121°C) – For sterilizing fabrics, Sterilize temperature: 250°F (121°C) – Sterilize time: 30 minutes – Dry time: 15 minutes (Note: A gravity cycle, adjusted to 270°F for 25-minute sterilize exposure time, can be used for processing fabric packs.)
270°F (132°) Gravity Cycle: For sterilizing hard goods. Sterilize temperature: 270°F (132°C) – Sterilize time: 15 minutes – Dry time: 30 minutes (Note: 270°F temperature must be set by the user)
Note: SFPP Cycle: For efficient, high-volume processing of porous loads. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization.
Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes
WRAP/SFPP Cycle: For efficient, high-volume processing of heat- and moisture-stable materials, such as double wrapped instrument trays. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization. – Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes
Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures. » Sterilize temperature: 250°F (121°C) – Factory programmed sterilize time: 45 minutes – Dry time: not applicable
275°F (135°C) Prevacuum Cycle: This cycle is used for efficient, high-volume processing of heat- and moisturestable materials in double-wrapped instrument trays. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. » Sterilize temperature: 275°F (135°C) » Sterilize time: 3 minutes » Dry time: 16 minutes
Utility Requirements
• 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp
Electrical–Vacuum Pump
• 208/240 Volt, 50/60 Hz, 3-phase, 6 Amp per phase.