Steris AMSCO Century V-148 Steam Sterilizer (148H and G-148)

Request Quote

The Steris AMSCO Century V-148 Steam Sterilizer (148H and G-148) is a fast, efficient pre-vacuum steam sterilizer designed for heat- and water-resistant loads, delivering the same capabilities as a gravity model with enhanced processing speed. It features a 26″ x 37.5″ x 48″ (767 L) chamber and cycles including pre-vacuum, gravity, flash, express, leak test, and daily air removal test. Highlights include a fluorescent display, ink-on-paper impact printer, emergency manual exhaust valve, a non-lubricated, steam-activated door seal, and helpful screens for programming and alarm troubleshooting. The unit offers broad PC communication compatibility, an electronic water conservation valve, and optional stainless steel side panels, with available pre-vacuum and gravity configurations to fit your application.

To submit a quote request for this product please fill out the form below.
Product Quote Request
Name
Name
Quantity *

Description

Steris AMSCO Century V-148 Steam Sterilizer (148H and G-148)

Steris V-148 Steam Sterilizers are designed for fast, efficient sterilization of heat and water-resistant materials with the same capabilities as a gravity sterilizer. The V148 pre-vacuum sterilizer is equipped with pre-vac, gravity, flash, express, leak test, and daily air removal test cycles. With a 26″ x 37.5″ x 48″ (767 Liters) chamber, the V148 sterilizer provides enough space fit all of your equipment and sterilize with speed and reputable quality in mind.

Features:

  • Florescent display for easy reading
  • Ink on paper impact printer
  • 26″ x 37.5″ x 48″ chamber (767 Liters)
  • Pre-vacuum Steam Sterilizer
  • Fast & efficient
  • Emergency manual exhaust valve
  • Non-lubricated, steam-activated door seal
  • Help screens for programming/troubleshooting alarm conditions
  • Communication interface compatibility w/ most PC devices
  • Electronic water conservation valve
  • Optional Stainless steel side panels available.

Additional Options:

  • Cabinet Package
  • Loading Cart
  • Integral Steam Generator
  • Water Filtration Options

Breakdown of Prevacuum vs. Gravity 

  • Gravity - designed for sterilization of liquids in flasks with vented closures, at 250°F (121°C), and also heat- and moisture-stabile goods at 270°F (132°C).
  • Prevacuum – designed for efficient, high-volume processing of heat- and moisture-stabile materials, such as fabrics, wrapped hard goods, glassware or animal bedding, at 270°F (132°C). Prevacuum units will also run gravity-type cycles.

Gravity units operate by the gravity downward displacement air-removal principle. Prevacuum sterilizers have a chamber evacuation system for high-speed, high-volume sterilization. Both types use steam-under-pressure as the sterilizing agent.

 

Prevacuum configuration sterilizers are factory programmed with the following cycles:

  • 270°F (132°C) Prevacuum (with either 5-minute or 20-minute dry phases) – For efficient, high volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 5 minutes or 20 minutes
  • Gravity, Liquid, 250°F (121°C) – For sterilizing fabrics, Sterilize temperature: 250°F (121°C) – Sterilize time: 30 minutes – Dry time: 15 minutes (Note: A gravity cycle, adjusted to 270°F for 25-minute sterilize exposure time, can be used for processing fabric packs.)
  • 270°F (132°) Gravity Cycle: For sterilizing hard goods. Sterilize temperature: 270°F (132°C) – Sterilize time: 15 minutes – Dry time: 30 minutes (Note: 270°F temperature must be set by the user)
    • SFPP configuration sterilizers are factory programmed with the following cycles: WRAP/SFPP, SFPP, Prevacuum, Gravity, Liquid and Test Cycles.
    • Note: SFPP Cycle: For efficient, high-volume processing of porous loads. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization.
    • Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes
  • WRAP/SFPP Cycle: For efficient, high-volume processing of heat- and moisture-stable materials, such as double wrapped instrument trays. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization. – Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes
  • Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures. » Sterilize temperature: 250°F (121°C) – Factory programmed sterilize time: 45 minutes – Dry time: not applicable
  • 275°F (135°C) Prevacuum Cycle: This cycle is used for efficient, high-volume processing of heat- and moisturestable materials in double-wrapped instrument trays. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. » Sterilize temperature: 275°F (135°C) » Sterilize time: 3 minutes » Dry time: 16 minutes

Utility Requirements:

  • Steam
    • 1” NPT
    • 50 to 80 psig (3.5 – 5.6 bar) dynamic
    • 97% – 100% vapor quality
  • Drain
    • 2” ODT drain terminal
  • Electrical – Controls
    • 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp
  • Electrical–Vacuum Pump
    • 208/240 Volt, 50/60 Hz, 3-phase, 6 Amp per phase.
    • 480 Volt, 50/60 Hz, 3-phase, 3 Amp per phase.
  • Sterilizer Feed Water
    • 1″ NPT, 20 to 50 psig (1.4 to 3.5 bar) dynamic. Water is used for vacuum pump, heat exchanger and trap cooling.