Steris Amsco Century V160

Request Quote

Steris V160 Steam Sterilizers are designed for fast, efficient sterilization of heat and water-resistant materials with the same capabilities as a gravity sterilizer. The V160 pre-vacuum sterilizer is equipped with pre-vac, gravity, flash, express, leak test, and daily air removal test cycles. With a 26″ x 37.5″ x 60″ chamber, the V160 sterilizer provides enough space fit all of your equipment and sterilize with speed and reputable quality in mind.

To submit a quote request for this product please fill out the form below.
Product Quote Request
Name
Name
Quantity *

SKU: AM-V160 Category: Tag:

Description

Steris V160 Steam Sterilizers are designed for fast, efficient sterilization of heat and water-resistant materials with the same capabilities as a gravity sterilizer. The V160 pre-vacuum sterilizer is equipped with pre-vac, gravity, flash, express, leak test, and daily air removal test cycles. With a 26″ x 37.5″ x 60″ chamber, the V160 sterilizer provides enough space fit all of your equipment and sterilize with speed and reputable quality in mind.

Features:
Florescent display for easy reading
Ink on paper impact printer
26″ x 37.5″ x 60″ chamber
Pre-vacuum Steam Sterilizer
Fast & efficient
Emergency manual exhaust valve
Non-lubricated, steam-activated door seal
Help screens for programming/troubleshooting alarm conditions
Communication interface compatibility w/ most PC devices
Electronic water conservation valve
Optional Stainless steel side panels available.

Additional Options: Cabinet Package, Loading Cart, Integral Steam Generator, Water Filtration Options

All processing cycles factory programmed into the sterilizer control have been validated to AAMI ST8.

Prevacuum configuration sterilizers are factory programmed with the following cycles:

270°F (132°C) Prevacuum (with either 5-minute or 20-minute dry phases) – For efficient, high volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 5 minutes or 20 minutes

Gravity, Liquid, 250°F (121°C) – For sterilizing fabrics, Sterilize temperature: 250°F (121°C) – Sterilize time: 30 minutes – Dry time: 15 minutes (Note: A gravity cycle, adjusted to 270°F for 25-minute sterilize exposure time, can be used for processing fabric packs.)

270°F (132°) Gravity Cycle: For sterilizing hard goods. Sterilize temperature: 270°F (132°C) – Sterilize time: 15 minutes – Dry time: 30 minutes (Note: 270°F temperature must be set by the user)

SFPP configuration sterilizers are factory programmed with the following cycles: WRAP/SFPP, SFPP, Prevacuum, Gravity, Liquid and Test Cycles.

Note: SFPP Cycle: For efficient, high-volume processing of porous loads. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization.
Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes

WRAP/SFPP Cycle: For efficient, high-volume processing of heat- and moisture-stable materials, such as double wrapped instrument trays. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization. – Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes

Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures. » Sterilize temperature: 250°F (121°C) – Factory programmed sterilize time: 45 minutes – Dry time: not applicable

275°F (135°C) Prevacuum Cycle: This cycle is used for efficient, high-volume processing of heat- and moisturestable materials in double-wrapped instrument trays. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. » Sterilize temperature: 275°F (135°C) » Sterilize time: 3 minutes » Dry time: 16 minutes

Utility Requirements

Steam
• 1” NPT
• 50 to 80 psig (3.5 – 5.6 bar) dynamic
• 97% – 100% vapor quality
Drain
• 2” ODT drain terminal
Electrical – Controls

• 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp

Electrical–Vacuum Pump

• 208/240 Volt, 50/60 Hz, 3-phase, 6 Amp per phase.

• 480 Volt, 50/60 Hz, 3-phase, 3 Amp per phase.
Sterilizer Feed Water
1″ NPT, 20 to 50 psig (1.4 to 3.5 bar) dynamic. Water is used for vacuum pump, heat exchanger and trap cooling.