Steris AMSCO SV-136 Scientific Medium Steam Sterilizer

Request Quote

The Steris AMSCO SV-136 Scientific Medium Steam Sterilizer is a high-capacity, pre-vacuum sterilizer designed for fast, reliable sterilization of heat- and moisture-stable materials. Featuring a spacious 26″ x 37.5″ x 36″ (575 liters) chamber, it provides ample room to sterilize all your laboratory equipment efficiently. Equipped with a user-friendly fluorescent display, ink-on-paper impact printer, and help screens for troubleshooting, it ensures ease of operation and thorough documentation. The SV-136 supports multiple cycles, including pre-vacuum, gravity, flash, express, leak test, and daily air removal, making it versatile for various sterilization needs. Built with a non-lubricated, steam-activated door seal and an emergency manual exhaust valve, it guarantees safety and performance. Optional stainless steel side panels are available for added durability. Designed for high-quality, repeatable performance, the SV-136 is ideal for sterilizing liquids, media, glassware, instruments, clothing, and laboratory waste in demanding scientific environments.

To submit a quote request for this product please fill out the form below.
Product Quote Request
Name
Name
Quantity *

SKU: AM-SV136 Category: Tags: , , , , Brand: ,

Description

Steris AMSCO SV-136 Scientific Medium Steam Sterilizer

Steris AMSCSO SV-136 Scientific Steam Sterilizers are designed for fast, efficient sterilization of heat and water-resistant materials with the same capabilities as a gravity sterilizer. The SV-136 pre-vacuum sterilizer is equipped with pre-vac, gravity, flash, express, leak test, and daily air removal test cycles. With a 26″ x 37.5″ x 36″ (575 Liters) chamber, the SV-136 sterilizer provides enough space fit all of your equipment and sterilize with speed and reputable quality in mind and designed for sterilization of nonporous heat- and moisture-stable goods, sterilization of liquids and media applications. This laboratory steam sterilizer provides high quality repeatable performance and accountability for a wide range of lab applications, including: liquids, media, glassware, instruments, clothing, waste, media clothing and laboratory items. 

Features:

  • Florescent display for easy reading
  • Ink on paper impact printer
  • 26″ x 37.5″ x 36″ chamber (575 L)
  • Pre-vacuum Steam Sterilizer
  • Fast & efficient
  • Emergency manual exhaust valve
  • Non-lubricated, steam-activated door seal
  • Help screens for programming/troubleshooting alarm conditions
  • Communication interface compatibility w/ most PC devices
  • Electronic water conservation valve
  • Optional Stainless steel side panels available.

Additional Options: Cabinet Package, Loading Cart, Integral Steam Generator, Water Filtration Options

Breakdown of Prevacuum vs. Gravity 

  • Gravity - designed for sterilization of liquids in flasks with vented closures, at 250°F (121°C), and also heat- and moisture-stabile goods at 270°F (132°C).
  • Prevacuum – designed for efficient, high-volume processing of heat- and moisture-stabile materials, such as fabrics, wrapped hard goods, glassware or animal bedding, at 270°F (132°C). Prevacuum units will also run gravity-type cycles.

Gravity units operate by the gravity downward displacement air-removal principle. Prevacuum sterilizers have a chamber evacuation system for high-speed, high-volume sterilization. Both types use steam-under-pressure as the sterilizing agent.

 

Interior Chamber Dimensions:

  • 26″ x 37.5″ x 36″ (660 x 950 x 910 mm)

Vacuum Pump Electric Service:

  • 208/240 VAC, 60 Hz, 3-Phase, 6A per phase
  • 480 VAC, 60 Hz, 3-Phase, 3A per phase

Prevacuum configuration sterilizers are factory programmed with the following cycles:

270°F (132°C) Prevacuum (with either 5-minute or 20-minute dry phases) – For efficient, high volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 5 minutes or 20 minutes

Gravity, Liquid, 250°F (121°C) – For sterilizing fabrics, Sterilize temperature: 250°F (121°C) – Sterilize time: 30 minutes – Dry time: 15 minutes (Note: A gravity cycle, adjusted to 270°F for 25-minute sterilize exposure time, can be used for processing fabric packs.)

270°F (132°) Gravity Cycle: For sterilizing hard goods. Sterilize temperature: 270°F (132°C) – Sterilize time: 15 minutes – Dry time: 30 minutes (Note: 270°F temperature must be set by the user)

SFPP configuration sterilizers are factory programmed with the following cycles: WRAP/SFPP, SFPP, Prevacuum, Gravity, Liquid and Test Cycles.

Note: SFPP Cycle: For efficient, high-volume processing of porous loads. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization.
Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes

WRAP/SFPP Cycle: For efficient, high-volume processing of heat- and moisture-stable materials, such as double wrapped instrument trays. This process incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization. – Sterilize temperature: 270°F (132°C) – Sterilize time: 4 minutes – Dry time: 20 minutes

Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures. » Sterilize temperature: 250°F (121°C) – Factory programmed sterilize time: 45 minutes – Dry time: not applicable

275°F (135°C) Prevacuum Cycle: This cycle is used for efficient, high-volume processing of heat- and moisturestable materials in double-wrapped instrument trays. This process incorporates a series of pressure/vacuum pulses to condition the load prior to sterilization. » Sterilize temperature: 275°F (135°C) » Sterilize time: 3 minutes » Dry time: 16 minutes

Utility Requirements

Steam

• 1” NPT
• 50 to 80 psig (3.5 – 5.6 bar) dynamic
• 97% – 100% vapor quality
Drain
• 2” ODT drain terminal
Electrical – Controls

• 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp

Electrical–Vacuum Pump

• 208/240 Volt, 50/60 Hz, 3-phase, 6 Amp per phase.

• 480 Volt, 50/60 Hz, 3-phase, 3 Amp per phase.
Sterilizer Feed Water
1″ NPT, 20 to 50 psig (1.4 to 3.5 bar) dynamic. Water is used for vacuum pump, heat exchanger and trap cooling.

Additional information