Zimmer ATS 750
The Zimmer A.T.S. 750 is a single-port automatic pneumatic surgical tourniquet system intended for qualified medical professionals to temporarily occlude blood flow in a patient’s extremities during extremity surgery.
Key features
- Uses ambient air — no nitrogen tank required.
- Electronic / microprocessor-regulated pressure for automatic pressure control.
- Single-port system designed for single-port, single-bladder cuffs.
- 30-minute battery backup for short-term mobility/backup operation.
- Audible and visual alarms for unsafe or fault conditions.
- Self-check calibration at power-up.
- Large red LED displays for pressure and time, plus a green AC power indicator.
- Tabletop or IV-pole mounting capability.
- Dual-cuff capability only with an optional Dual Control Valve accessory for bilateral/Bier Block/IVRA use.
Main specifications
| Category | Spec |
|---|---|
| Pressure range | 50–475 mmHg, adjustable in 5 mmHg increments |
| Pressure accuracy | ±5 mmHg over 50–475 mmHg |
| Pressure regulation | ±6 mmHg of set point, 10-second average, under non-transient conditions without external leaks |
| Maximum cuff pressure | 475 mmHg |
| Time alarm range | 5–240 minutes, 5-minute increments per service manual |
| Timer accuracy | 0.25% of elapsed time |
| Power input | 100–240 VAC, 50/60 Hz, auto-switching |
| Line current | 670 mA RMS @ 120 V |
| Input power | 53 W typical |
| Battery | Rechargeable 12 V sealed lead-acid, 1.2 Ah |
| Battery run time | Minimum 30 minutes with fully charged batteries |
| Battery recharge time | 24 hours |
| Power cord | Type SJT, AWG 16, 14 ft / 4.27 m, hospital-grade 3-prong plug |
| Fuses | Two time-delay 1.0 A, 250 V fuses |
| Dimensions | 8 in H × 7 in W × 7.9 in D / 20.3 × 17.8 × 20.1 cm |
| Weight | 7.7 lb / 3.5 kg; |
| Displays | Pressure: red 14-segment LED; Time: red 14-segment LED; AC indicator: green LED |
| Controls | On/Standby, rotary knob, Pressure, Time, Inflate, Deflate, Alarm Silence |
| Safety classification | UL 60601-1 Class I and internally powered, Type B, continuous operation; not for use with flammable anesthetic or gases |
| Internal diagnostics | Program, memory, watchdog timer, transducer calibration, improper valve actuation |
Alarm / safety functions
The system provides alarms for low cuff pressure, high cuff pressure, cuff-side leak, time exceeded, cuff inflated on power-up, cuff not deflated, and line occlusion. Low/high pressure alarms trigger at about 15 mmHg below/above set point, and leak alarm indicates a substantial leak lasting more than 7 seconds. Most audible alarms can be silenced for 30 seconds using the Alarm Silence switch.