Description
HC-900 26 x 37-1/2 x 42″ Chamber
AMSCO® Evolution® Steam Sterilizers are configured for prevacuum sterilization of heat- and moisture-stable materials used in healthcare facilities. These units can be equipped with prevacuum, Steam Flush Pressure Pulse (SFPP), gravity, liquid, and daily air removal test cycles which use, for example, a Bowie-Dick test pack or a Dart® testing apparatus to determine the effectiveness of removing air from a sterilizer chamber.
Hinged door for 26 x 37.5″ (660 x 953 mm) sterilizers. Door may be hinged on either left side or right side.
All plumbing components are mounted to a free-standing, modular rack (stand). The stand connects to the core sterilizer assembly during installation
Able to fit four full shelves of 25 lb
Facility Steam Size: Copper/Brass 3/4″ NPT, Stainless Steel 1″ Sanitary TRI-CLAMP®* fitting.
Pressure: 50 to 80 psig (3.45 to 5.5 bar), dynamic, condensate free, and 97 to 100% vapor quality
Average Steam consumption: 63lbs/hr Steam
Peak Steam Consumption: 310 lbs/hr Steam