Description
Generator:
8.0 kW, 100 mA, 125 kVp (see power requirements below)
High-Frequency Ripple: 300 kHz
40 to 125 kVp in 1 kVp steps
Touch Screen
Allows customization of 6 APR regions (skull, thorax, spine, abdomen/pelvis, lower and upper extremities)
Small, medium, large, and pediatric patient sizes
User-friendly multi-language selection
14 mA stations: 5, 6.4, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80 and 100
0.001-10 second exposure time range, 41 total in 25% steps
0.1 to 250 mAs, 34 total in 25% steps
X-Ray Tube:
0.6mm x 2.8mm dual focus stationary anode tube, 40,000HU/15 Degree
Collimator:
The collimator, LASER, certified manua
Mechanics:
Stand on base with wheels for easy maneuverability
The gas spring arm supports the tube to accommodate easy positioning
Positive mechanical brakes
The Cassette storage box accommodates up to 8 cassettes
Inclinometer
Specifications:
Vertical tube travel, 62″ (16.7″-78.7″)
118 lbs total weight
Power:
220 – 240 VAC; Requires solid 20 amp breaker for full pawer (can be de-rated)
Power Cord, 19.7′ long (6 meters)
Exposure Cord, 9.8′ long (3 meters)
Single-phase line Automatic Regulation, Automatic Output Power Adaptation to Power Line Conditions